M Series Service Manual

11

4.0 Power Supply Test (Optional)

Tools Needed:

• 1 red miniature alligator to miniature alligator test lead.

• 1 black miniature alligator to miniature alligator test lead.

• DC power supply (15 Amp minimum).

• 0.1 ohm



resistor (

¼

W or greater).

• 1000 ohm 1% resistor (

¼

W or greater).

• Fluke 75 multimeter or equivalent.

Test Setup:

Make sure the unit and power supply are turned off.

Connect one end of the black lead to the “-” terminal in the battery well.

Connect the other end of the black lead to the “-” terminal of the power supply.

Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it.
Connect the other end of the red lead to the “+” terminal of the power supply.

Set the power supply voltage to 7V.

CAUTION

Be sure to connect the power supply properly to the M Series battery well terminals or damage to the unit may result. Do NOT raise

the power supply voltage above 12V.