M Series Service Manual

67

NO SHOCK ADV

No shock advised. Advisory message when analysis

finds non-shockable rhythm.

NOISY ECG

Number of noisy analysis intervals exceeds

threshold.

Stop all patient

movement. Check

connections. Press

Analyze button again.

OPEN AIR DISCHARGE

Cap voltage too high after discharge attempt, e.g.,

full energy discharge did not occur.

Replace paddles, and,

or high voltage module

and system board.

PACER DISABLED

User prompt issued simultaneously with other pace

faults if pacing is disabled.

Replace high voltage

module or system

board.

PACER FAULT 115

Flyback pulse width control circuit is not under

proper control of the processor and gate array.

Replace high voltage

module, capacitor, or

system board.

PACER FAULT 116

Failure to detect XPACE_ON.

Replace high voltage

module, capacitor, and/

or system board.

PACER FAULT 117

Pace relay is stuck closed.

Replace high voltage

module, capacitor, and/

or system board.

PACER FAULT 121

During pace, the pace pulse width <30ms or >50ms.

Replace high voltage

module, or system

board.

PACER FAULT 122

Pace current is more than 15mA above and below

selected value.

Replace high voltage

module, or system

board.

Error Message

Explanation

User
Advisory

Technical Action