M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 76

Advertising
background image

M Series Service Manual

66

ECG V6 LEAD OFF

Chest lead V6 is not properly attached to patient.

Reattach V lead. Check

cable.

ENTER ACCESS CODE

Manual mode access code needed.

Enter access code to

enter manual mode

with AED.

ERASING REPORT

Summary report being erased.

ECU CRC FAULT

Invalid ECG samples detected over a one second

period.

Turn off unit and then

turn on to reset. If fault

persists, replace system

board.

EtCO

2

COM ERROR

No or invalid communication from EtCO

2

module.

Return unit for service

to ZOLL Technical

Service Department.

FAX DIALING

Preparation for sending fax.

FAX DONE

Transmission complete.

FAX PREPARING

Preparing fax for transmission.

FAX SENDING

Transmitting fax.

INSERT CARD

Check memory card

No card installed in unit during manual or semi-

automated modes.

LOW BATTERY

Low battery.

Replace battery or plug

into AC power. Replace

charger.

NO QRS DETECT

Unit is in sync mode and heart rate is < 20 BPM or

QRS amplitude is too low for proper

synchronization.

Increase ECG size and/

or change lead.

Error Message

Explanation

User
Advisory

Technical Action