M Series Service Manual

45

21.4

On the M Series unit, press the
Set Low
softkey to calibrate the
transducer to a 0 mmHg pressure
reading.

The NIBP pressure transducer registers its voltage
output at a known pressure of 0 mmHg. The field
adjacent to the 0 mmHg value changes to PASS.

Note:

If the M Series unit displays a FAIL reading,
verify the NIBP simulator’s pressure setting and
connection to the M Series and repeat the step.

o

o

21.5

On the NIBP Simulator, set the
pressure parameter to
250 mmHg, and then press Start
on the NIBP simulator.

The NIBP Simulator displays a pressure reading of 250
mmHg.

o

o

21.6

On the M Series unit, press the
Set High
softkey to calibrate the
transducer to a 250 mmHg
pressure reading.

The NIBP pressure transducer registers its voltage
output at a known pressure of 250 mmHg. The field
adjacent to the 250 mmHg value changes to PASS.

Note:

If the M Series unit displays a FAIL reading,
verify the NIBP Analyzer’s pressure setting and
connection to the M Series and repeat the step.

o

o

21.7

On the NIBP simulator, set the
pressure parameter to stimulate a
different cuff pressure (for
example, 205 mmHg).

The NIBP simulator displays the specified pressure
reading.

o

o

21.8

On the M Series unit, press the
Read Cuff
softkey when the
simulator’s Measured screen
reaches 205.

The value that the M Series unit displays is accurate
within +/- 3mmHg of the pressure parameter value set on
the NIBP simulator.

o

o

21.9

On the M Series unit, press the
EXIT softkey twice.

The M Series returns to the main NIBP Service Mode
screen, then to normal Monitor mode operation.

o

o

Record your results on the Maintenance Tests Checklist.

Do this...

Verify that...

Pass/
Fail