Removing the upper housing assembly – ZOLL M Series Defibrillator Rev R User Manual

Page 96

Advertising
background image

M Series Service Manual

86

3. Removing

the Upper
Housing
Assembly

Tool Required

• No. 1 and 2 Phillips screwdriver

Set-up

Remove Front Panel (See Step 2.)

1. Remove the two screws securing the

universal cable.

2. Do not lose the O-ring when removing the

universal cable.

3. Remove the three Phillips screws in the

front of the Upper Housing.

To reinstall the Upper Housing Assembly,
reverse the above steps.

Ensure the battery pin gaskets are properly set.

Step 1: Remove two screws from the back side of
the Upper Housing Assembly and three screws
from the front.

Step 2: Secure lower housing to the table by
pressing downward in the paddle well. Using the
carry handle, lift the Upper Housing upward.

Step 3: Make sure that the rubber gaskets are
covering the battery contact pins. If the gaskets
are still seated in the housing, remove and place
them onto the contact pins. Before installing the
Upper Housing ensure that the battery pin gaskets
are set properly.

battery pin gasket