M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 37

Advertising
background image

M Series Service Manual

27

13.8

Press the ENERGY SELECT up
arrow until the unit displays 100J.

The unit displays the message, DEFIB 100J SEL.

o

o

o

13.9

Press the CHARGE button. Wait
for the SHOCK button to
illuminate.

The unit displays the message, DEFIB 100J RDY.

o

o

o

13.10

Press the SHOCK button.

The unit discharges 83J-122J (monophasic 85J-115J)
into the simulator.

o

o

o

13.11

Press the ENERGY SELECT up
arrow until the unit displays 200J.

The unit displays the message, DEFIB 200J SEL.

o

o

o

13.12

Press the CHARGE button. Wait
for the SHOCK button to
illuminate.

The unit displays the message, DEFIB 200J RDY.

o

o

o

13.13

Press the SHOCK button.

The unit discharges 182J-247J (monophasic 170J-
230J) into the simulator.

o

o

o

13.14

(Monophasic Units) Press the
ENERGY SELECT up arrow unit
the unit displays 300J.

The unit displays the message, DEFIB 300J SEL.

o

o

o

13.15

Press the CHARGE button. Wait
for the SHOCK button to
illuminate.

The unit displays the message, DEFIB 300J RDY.

o

o

o

13.16

Press the SHOCK button.

The unit discharges 255J-345J into the simulator.

o

o

o

13.17

(Monophasic Units) Press the
ENERGY SELECT up arrow unit
the unit displays 360J.

The unit displays the message, DEFIB 360J SEL.

o

o

o

13.18

Press the CHARGE button. Wait
for the SHOCK button to
illuminate.

The unit displays the message, DEFIB 360J RDY.

o

o

o

13.19

Press the SHOCK button.

The unit discharges 306J-414J into the simulator.

o

o

o

Do this...

Verify that...

Pass/Fail/
NA

Advertising