M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 75

Advertising
background image

M Series Service Manual

65

ECG FAULT 4

Communication fault between ECG processor and

main processor.

Turn off unit and then

turn on to reset. If fault

persists, replace system

board.

ECG FAULT 5

ECU RAM test failure, or ROM checksum test

failure.

Turn off unit and then

turn on to reset. If fault

persists, replace system

board.

ECG LEAD OFF

One or more ECG leads are not properly connected

when leads are selected as input.

Check cable and patient

connection. Change

electrodes. Prepare

patient’s skin.

ECG TOO LARGE

ECG signal too large for accurate shockable rhythm

analysis.

Reduce ECG size.

ECG TOO SMALL

ECG signal too small for accurate shockable rhythm

analysis.

Increase ECG size.

ECG V1 LEAD OFF

Chest lead V1 is not properly attached to patient.

Reattach V lead. Check

cable.

ECG V2 LEAD OFF

Chest lead V2 is not properly attached to patient.

Reattach V lead.

Check cable.

ECG V3 LEAD OFF

Chest lead V3 is not properly attached to patient.

Reattach V lead.

Check cable.

ECG V4 LEAD OFF

Chest lead V4 is not properly attached to patient.

Reattach V lead.

Check cable.

ECG V5 LEAD OFF

Chest lead V5 is not properly attached to patient.

Reattach V lead.

Check cable.

Error Message

Explanation

User
Advisory

Technical Action