M Series Service Manual

34

17.4

Using the Index SpO

2

Simulator,

press the BPM+ or BPM- softkey
until the heart rate is 230 BPM.

The SpO

2

simulator screen displays an SpO

2

rate of 230

BPM.

Note that you may need to wait up to 2 minutes for the
information to appear on the ZOLL display.

The M Series monitor displays an SpO

2

saturation of 96-

100%.

The M Series monitor displays a heart

rate of 226-234

BPM.

o

o

17.5

Using the Index SpO

2

Simulator,

press the BPM- softkey until the
heart rate is 50 BPM.

The SpO

2

simulator screen displays an SpO

2

saturation

of 96-100%.

The M Series monitor displays a heart rate of 46-54
BPM.

o

o

o

o

17.6

Using the Index SpO

2

Simulator,

press the 02+ softkey until the
SpO

2

output is at 72%.

The SpO

2

simulator screen displays an SpO

2

saturation

of 70-74%.

The M Series monitor displays a heart rate of 46-54
BPM.

o

o

o

o

17.7

Press Wave 2 softkey.

Select the SpO2 waveform.

For CCT units: Press the Traces
softkey, and then select SET
TRACE 2
, and press Enter.
Select
SPO2, press Enter.

Plethysmographic waveform appears on the ZOLL
display.

o

o

17.8

Press RECORDER.

The plethysmographic waveform prints on the strip chart
paper.

o

o

17.9

Using the Index SpO

2

Simulator,

press the BPM- softkey until the
heart rate is at 230 BPM.

The SpO

2

simulator screen displays an SpO

2

saturation

rate of 70-74%.

The M Series monitor displays a heart rate in the heart
position of
226-234 BPM.

o

o

17.10

Remove the Masimo

patient

cable.

Do this...

Verify that...

Pass/
Fail