M Series Service Manual

68

PACER FAULT 123

Measured pace rate is too fast compared to selected

rate.

Replace high voltage

module or system

board.

PACER FAULT 126

Issued in conjunction with message 122. Pace

current is more than 15mA and below selected value.

Replace high voltage

module.

PADDLE FAULT

Cannot detect type of accessory attached to the

universal cable.

Replace paddles,

internal paddles, system

board, high voltage

module and/or

universal cable.

PERFORM CPR

Advisory message in AED auto defib mode.

PLACE ON ZERO CELL

EtCO

2

sensor cable plugged into unit for the first

time. Zeroing error or probe drift error detected.

Zero sensor. Replace

sensor. Return to ZOLL

for service.

POOR LEAD CONTACT

One or more ECG leads are poorly connected or not

connected to patient. (User configurable.)

Check electrode

attachment to patient,

cable connector to

electrode, cable to unit

connector. Broken unit.

POOR PAD CONTACT

Electrode impedance exceeds threshold.

Ensure pads are

coupled to patient.

Check /replace pads or

universal cable.

Check impedance

circuit calibration.

Replace system board.

PRESS ANALYZE

Alternate message for check patient prompt.

Error Message

Explanation

User
Advisory

Technical Action