M Series Service Manual

59

AUDIO FAULT 136

Audio DSP hardware error.

Replace audio board.

Replace system

board.Turn unit off and

back on again.

AUDIO NOT

RECORDING

Audio is not recording.

Install PCMCIA card.

Replace system board.

AUDIO QUEUE FULL

Indicates that the audio output queue is full.

Additional voice prompts can't be queued at this

time.

None.

BATT HIGH CURRENT

Battery is charged and battery current is >.1 A or:

Battery is not charged and battery current is > 1.6 A.

Unplug from A/C.

Remove the battery for

20 seconds. Reconnect

all above. If the

problem persists,

replace battery and or

charger.

BATT HIGH VOLTAGE

Battery voltage > 15.5 v.

Replace battery and or

charger

BATT LOW CURRENT

Battery is not charged and battery current is <.35 A.

Replace battery and or

charger.

BATT LOW VOLTAGE

Battery voltage < 9.5 v.

Replace battery and or

charger.

BATT OVERCHARGE

Charger on for > 4 hours.

Replace battery and or

charger.

Error Message

Explanation

User
Advisory

Technical Action