Flash or arcing under

defibrillator pad.

Avoiding using alcohol and betadine in and around
the treatment area because these skin preparations
may lead to increased conductivity and/or bonding
between the electrode’s adhesive and skin.

Check for gel droop. If the gel has leaked out of the
gel treatment area, replace the electrode.

Ensure pads are coupling to the patient’s skin and
connected to the universal cable.

Check for dried out gel on the defibrillator pad.

Clip patient’s excessive hair. Do not shave hair.

Check expiration date. Replace pad if date has
expired.

Do not conduct chest compression through the
pads because the pads could be damaged leading to
the possibility of arcing and skin burns.

Apply the back electrode first. If the front electrode
is already in place when the patient is being
maneuvered for placement on the back, the front
may become partially lifted, possibly causing
arching and skin burns.

Ensure that wet gel pads are stored
flat.

Displayed HR not accurate.

No artifact present.

Verify heart rate flashes with each QRS on display.

Change lead selection.

Change ECG size.

Reposition ECG electrodes.

Reported Problem

Recommended User Action

Recommended Technical Action