M Series Service Manual

42

20.26

On the Patient Simulator, set BP
channel 1 to a static pressure of 20
mmHg (waveform 11) and press
EXECUTE.

20.27

Press the RECORDER button.

Verify that there is no deviation from the baseline greater
than 1 mm for the waveforms printed on stripchart and the
unit displays the waveforms for 20mmHg on the monitor
screen, and prints a stripchart of the traces.

o

o

o

20.28

Set the input and output resistance
to 350 ohms on the Transducer
Simulator, and set the gain to x10.

Connect the Transducer Simulator
cable to the P1 connector on the
unit and on the Pressure
Transducer Simulator, set the
pressure polarity to positive, and
select a pressure of 200 mmHg.

20.29

Zero pressure channel 1.

Verify the unit calibrates pressure channel 1 against a
positive pressure of 200 mmHg and displays the P1
ZEROED message.

o

o

o

20.30

On the Pressure Transducer
Simulator, set the pressure polarity
to positive, and select a pressure
of 500 mmHg.

Verify that the value displayed for the P1 pressure
channel is in the range of 291 - 300 mmHg. (Note: value
will be displayed within 15 seconds.)

o

o

o

20.31

Connect the simulator cable to the
P2 connector on the unit. On the
Pressure Transducer Simulator, set
the pressure polarity to positive,
and select a pressure of 200
mmHg.

20.32

Zero ART channel.

Verify the unit calibrates pressure channel 1 against a
positive pressure of 200 mmHg and displays the ART
ZEROED message.

o

o

o

Do this...

Verify that...

Pass/Fail/
NA