M Series Service Manual

24

NOTE: The universal cable should not be connected to any equipment at the beginning of this test.

Do this...

Verify that...

Pass/
Fail

11.1

Turn the selector switch to DEFIB
mode.

(For AED units, turn the selector
switch to ON and select Manual
mode.)

Set leads to PADS.

The unit displays the CHECK PADS/POOR PAD
CONTACT
message.

o

o

11.2

Connect the universal cable to the
MFC test port.

DEFIB PAD SHORT message displays.

o

o

11.3

Select energy level of 100J and
press the CHARGE button.

The charge time is greater than 2 seconds and less than
10 seconds and the unit displays the message, SELECT
30J FOR TEST
.

o

o

11.4

Press the SHOCK button.

The unit does not discharge. The unit displays the
message, DEFIB PAD SHORT.

o

o

11.5

Set energy level to 30J.

The unit discharges internally.

o

o

11.6

Press the CHARGE button.

The unit charges to 30J and displays message, DEFIB
30J READY.
The charge ready tone sounds.

o

o

11.7

Press and hold SHOCK button.

The unit discharges and displays the number of joules
delivered and the TEST OK message. For example, the
message at the top of the printed strip chart would read
as follows: 30 JOULES TEST OK. TEST_CUR=10-14A
DEFIB_IMPED=0.

The impedance value may range from 0 to 5 ohms.

If configured, the recorder prints a strip chart.

o

o

Record your results on the Maintenance Tests Checklist.