M Series Service Manual

63

DEFIB FAULT 72

General defib error.

Turn the unit to OFF

and back on. If fault

persists, replace high

voltage module.

DEFIB FAULT 76

Capacitor voltage too high for selected energy.

Replace high voltage

module or capacitor.

DEFIB FAULT 77

Capacitor voltage > than absolute rated max.

Replace high voltage

module or capacitor.

DEFIB FAULT 78

Unable to charge defib cap.

Replace high voltage

module or capacitor.

DEFIB FAULT 79

Defibrillator charging too slowly.

Replace high voltage

module or capacitor.

DEFIB FAULT 80

4 defibrillator faults detected within 20 second

period.

Replace high voltage

module or capacitor.

DEFIB FAULT 81

Discharge switch in undefined state.

Replace high voltage

module or capacitor.

DEFIB FAULT 84

“Upper” discharge transistor shorted (measured via

applicable A/D channel).

Replace high voltage

module.

DEFIB FAULT 85

“Lower” discharge transistor shorted (measured via

applicable A/D channel).

Replace high voltage

module.

DEFIB FAULT 86

One discharge switch closed during power up test.

Replace paddles,

control board or system

board.

DEFIB FAULT 87

Both discharge switches closed during power up test.

Replace paddles,

control board or system

board.

Error Message

Explanation

User
Advisory

Technical Action