M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 72

Advertising
background image

M Series Service Manual

62

CHECK SPO

2

SITE

Low or no perfusion in monitored finger or toe.

CHECK SPO

2

SENSOR

Reposition SpO

2

sensor on patient.

CLOCK FAULT 11

Real time clock oscillator failure.

Replace system board.

CLOCK FAULT 12

Real time clock back-up power supply failure. Found

oscillator stopped at power-up, but oscillator now

running when the system is running. (Oscillator only

runs when main power is applied).

Replace system board.

CLOCK FAULT 13

One of the set time units (seconds, minutes, year,

etc.) is out of range.

Replace system board.

CO

2

COMM ERROR

No or invalid communication from the EtCO

2

module.

Replace EtCO

2

module

and or system board.

CO

2

SENSOR WARM UP

EtCO

2

Sensor warming up.

Wait for sensor to warm

up. This process takes

up to approximately

one minute.

CONFIRM MANUAL

MODE

Displayed when manual mode is entered. Alerts user

to confirm that manual mode is desired.

DEFIB DISABLED

User prompt issued simultaneously with other faults

if defib is disabled.

Possible configuration

problem. Replace high

voltage module. Call

ZOLL Technical

Support.

DEFIB FAULT 71

More than 50 internal dumps occurred in less than 20

minutes.

Turn the unit to OFF

and back on. If fault

persists, replace high

voltage module.

Error Message

Explanation

User
Advisory

Technical Action