M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 136

Advertising
background image

M Series Service Manual

126

FET_MON

Analog voltage monitors the

condition of the discharge

transistors.

Provides a signature

voltage in case of a fault.

PAT_CUR

Bound on the range of 0 - 2.5

V and accommodates defib

currents of -50 to +100 A

Analog signal representing

the patient current during a

defib discharge.

L

OGIC

C

ONTROL

S

IGNAL

O

PERATION

XPWR_ENABLE: This

logic

signal from the gate array enables the charging circuit

when true, and inhibits the charging circuit when false.

SAFE:

This logic control signal is generated by the Main System Board

to halt the pace/defib function in the event of a detected fault.

SHUTDOWN:

This logic signal is true during reset and fault conditions. (VCC

error, watchdog error, etc.) and

halts operation of the PD

generator.

XPACE_ON:

This logic signal is generated by an optocoupler, and indicates

that the pace output circuit is active. It is '0' true when pace

current is flowing.

XSAFREL:

Logic signal from the GA that operates the safety relay when '0'

true.

XPACEREL:

Logic signal from the GA that operates the pace relay when '0'

true. XPATREL: Logic signal from the GA that operates the solid

state patient relay when '0' true.

XPAT_ENABLE:

Logic signal from the processor controlled by the monitor that

grants operation of the solid state patient relay. It is false during

pacing.

A

NALOG

V

OLTAGE

O

PERATION

C

OMMENT