M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 32

Advertising
background image

M Series Service Manual

22

10.9

Remove an EEG lead wire from
the QED 6 analyzer.

The alarm symbol flashes and the heart symbol stops
flashing. The ECG LEAD OFF alarm tone sounds.
Recorder prints a stripchart showing a low heart rate, if
enabled.

o

o

o

10.10

Reattach ECG Lead wire to the
QED 6 Analyzer and hold the
ALARM SUSPEND button on
unit for 4 seconds.

The unit displays the alarm symbol with an “X” through
it.

The heart symbol flashes with each QRS wave.

o

o

o

10.11

Press the ALARM SUSPEND
button to enable the alarms.

The alarm symbol does not have an “X” through it.

o

o

o

10.12

Set QED 6 Analyzer to 160 BPM
or higher.

The Heart Rate Value is highlighted, the alarm tone
sounds, and the alarm and the heart symbol both flash.

o

o

o

10.13

Press the ALARM SUSPEND
button on the M Series unit.

The alarm is suspended for 90 seconds. The unit
displays the alarm symbol with an “X” through it, and
the heart symbol flashes with each QRS wave.

Note: Software version 35.00-36.XX, the alarm is
suspended until the alarm button is press again, or the
fault is cleared.

o

o

o

10.14

Press and hold ALARM
SUSPEND
for 4 seconds to
disable alarms.

o

o

o

Record your results on the Maintenance Tests Checklist.

Do this...

Verify that...

Pass Fail/
NA