M Series Service Manual

73

USE PADS

(AUTO DEFIB MODE ONLY.)

Attempt to defib with paddles in auto defib (AED)

mode. Defib only allowed using PADS in AED

modes.

USE PADS TO PACE

External paddles detected in pace mode.

USE ROOM AIR

ADAPTER

Adapter zeroing started with EtCO

2

in the adapter or

the adapter is on the REF or “0” cell.

Place CO

2

sensor on

adapter in room air.

USER SETUP REQ

Both copies of stored cal/config data are bad or have

never been programmed.

Perform configuration

setup.

VF ALARMS OFF

VF alarms disabled in pace mode or when paddles

are selected as leads.

VX LEADS OFF

V lead not properly attached to patient. “X” denotes

lead number.

Reattach V lead.

ZERO CO

2

SENSOR

New EtCO

2

sensor needs to be zero calibrated.

Zero EtCO

2

sensor.

ZERO CO

2

ADAPTER

New EtCO

2

airway adapter needs to be zero

calibrated.

Zero EtCO

2

adapter.

Error Message

Explanation

User
Advisory

Technical Action