M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 70

M Series Service Manual
60
BRIDGE SHORT
Current higher than expected was detected during the
Biphasic bridge test or immediately following a
discharge.
Ensure pads/paddles
are used properly.
Attempt to clear the
message by turning the
Selector switch to off
then back to the desired
operating modes.
Replace bridge or high
voltage module.
BRIDGE TEST FAILED
Biphasic module not operating properly while
charging.
Charge again.
Attempt to clear the
message by turning the
Selector switch to OFF,
then back to the desired
operating mode.
Replace bridge or high
voltage module.
CABLE FAULT
(Auto defib mode only.)
Incorrect A/D reading for paddle ID (similar to
PADDLE FAULT).
Replace paddle set,
universal cable and/or
system board.
CANNOT CHARGE
Cannot charge when charge button pressed.
Replace high voltage
module or capacitor.
REPLACE CARD
Write errors during manual or semi-automated
modes.
May have configuration
card installed or write
protection on.
Error Message
Explanation
User
Advisory
Technical Action