M Series Service Manual

60

BRIDGE SHORT

Current higher than expected was detected during the

Biphasic bridge test or immediately following a

discharge.

Ensure pads/paddles

are used properly.

Attempt to clear the

message by turning the

Selector switch to off

then back to the desired

operating modes.

Replace bridge or high

voltage module.

BRIDGE TEST FAILED

Biphasic module not operating properly while

charging.

Charge again.

Attempt to clear the

message by turning the

Selector switch to OFF,

then back to the desired

operating mode.

Replace bridge or high

voltage module.

CABLE FAULT

(Auto defib mode only.)

Incorrect A/D reading for paddle ID (similar to

PADDLE FAULT).

Replace paddle set,

universal cable and/or

system board.

CANNOT CHARGE

Cannot charge when charge button pressed.

Replace high voltage

module or capacitor.

REPLACE CARD

Write errors during manual or semi-automated

modes.

May have configuration

card installed or write

protection on.

Error Message

Explanation

User
Advisory

Technical Action