M Series Service Manual

41

20.12

Zero the IBP channels (see step
20.9).

20.13

Press the Traces softkey.

Verify the unit displays the Traces menu.

o

o

o

20.14

Select the Set Trace 2 option and
press the Enter softkey.

Verify the M Series CCT unit displays the Trace Options
menu.

o

o

o

20.15

Verify the M Series CCT unit
displays the Trace Options menu.

Verify the unit assigns the P1 waveform to the second
trace display area and returns to the main menu.

o

o

o

20.16

Press the Traces softkey.

Verify the unit displays the Traces menu.

o

o

o

20.17

Select the Set Trace 3 option and
press the Enter softkey.

Verify the unit displays the Trace Options menu.

o

o

o

20.18

Select the ART option and press
the Enter softkey.

Verify the unit assigns the ART waveform to the third trace
display area and returns to the main menu.

o

o

o

20.19

Go to the IBP menu (press the
Param softkey and select the IBP
option).

Verify the unit displays the IBP menu.

o

o

o

20.20

Press the Range softkey.

Verify the unit displays the IBP Range menu.

o

o

o

20.21

Select the P1 Range option and
press the Enter softkey.

Verify the unit displays the Range Options menu.

o

o

o

20.22

Select the 0-30 mmHg option and
press the Enter softkey.

Verify the unit displays the P1 waveform on a 0-30 mmHg
scale and returns to the IBP menu.

o

o

o

20.23

Press the Range softkey.

Verify the unit displays the IBP Range menu.

o

o

o

20.24

Select the ART option and press
the Enter softkey.

Verify the unit displays the Range Options menu.

o

o

o

20.25

Select the 0-30 mmHg option and
press the Enter softkey.

Verify the unit displays the ART waveform on a 0-30
mmHg scale and returns to the IBP menu.

o

o

o

Do this...

Verify that...

Pass/Fail/
NA