Removing the high voltage module assembly – ZOLL M Series Defibrillator Rev R User Manual

Page 104

Advertising
background image

M Series Service Manual

94

7. Removing

the High
Voltage
Module
Assembly

Tools Required

• #2 Phillips screwdriver

• Exacto-Knife

WARNING! This unit may contain lethal voltages.
You must completely discharge the high voltage
capacitor by changing the energy selection during
the charge cycle of the defibrillator. Power off unit.
Wait three minutes before disassembly.

DO NOT SHORT THE TERMINAL ENDS OF
THE CAPACITOR. Refer to Section 8, Step 2.

Setup

1. Remove the Front Panel Assembly.

2. Remove the Upper Housing Assembly.

3. Remove the System Board.

4. Remove the Battery Interconnect Board.

5. Remove the High Voltage/Charger Assembly.

To reinstall the High Voltage Module Assembly, reverse the steps.