M Series Service Manual

64

DEFIB FAULT 94

Processor fault causing safety monitor port to be

non-functional.

Replace system board,

high voltage module or

capacitor.

DEFIB FAULT 95

Safe or shutdown line is not functional.

Replace high voltage

module.

DEFIB FAULT 96

XPATREL or XPAT_ENABLE is faulted or one of

the discharge transistors has shorted.

Replace high voltage

module.

DEFIB FAULT 108

VMON voltage is less than the target energy during

charging.

Replace high voltage

module or capacitor.

DEFIB FAULT 109

Defib capacitor voltage is greater than selected

energy when defibrillator is charging or ready.

Replace high voltage

module or capacitor.

DEFIB FAULT 111

Defib capacitor voltage has exceeded the absolute

maximum acceptable voltage.

Replace high voltage

module, capacitor, and

or system board.

DEFIB NOT CHARGED

Discharge button is pressed but the unit is not

charged.

DEFIB PAD SHORT

Measured impedance between high voltage leads of

MFC.

Ensure pads are

coupled to patient.

Check /replace pads or

universal cable.

Replace system board.

DISABLE SYNC

Sync mode active when analyze pressed in defib.

DISCHARGE FAULT

Defib capacitor voltage is not decreasing.

Replace high voltage

module, capacitor, and/

or system board.

Error Message

Explanation

User
Advisory

Technical Action