M Series Service Manual

71

SET PACE RATE

Multiple copy errors are the product of intended

software or memory errors. Multiple copies of pace

rate don't match. If error persists, unit could be

broken

Set pace rate. If broken,

replace system board.

SHOCK ADVISED

Advisory message when analysis finds a shockable

rhythm.

SpO

2

AMBIENT LIGHT

Ambient light is too bright.

Shield sensor from

ambient light. Replace

Sp0

2

sensor. Replace

Sp0

2

module

SpO

2

COMM ERR

No transmissions from SpO

2

unit received.

Communication error or no communication from

Sp0

2

module.

Replace Sp0

2

module

and/or system board.

SpO

2

PULSE SEARCH

Pulse search in progress.

SpO

2

SENSOR FAULT 1

Defective sensor.

Replace Sp0

2

sensor.

SpO

2

SENSOR FAULT 5

Unrecognized sensor.

Replace Sp0

2

sensor.

STAND CLEAR

(Auto defib mode only.) Single analysis mode just

turned on and defib idle. Patient rhythm is being

analyzed.

SYSTEM FAULT 2

MCU ROM checksum test failure or MCU RAM test

failure.

Turn off unit and then

turn on to reset. If fault

persists, replace system

board.

Error Message

Explanation

User
Advisory

Technical Action