0 leakage current test, M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 25
Advertising
M Series Service Manual
15
5.0 Leakage Current Test
Tools Needed:
See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Test Setup:
See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with
accessories: MFC, external paddles, and anterior/posterior paddles.
Table 1:
Maximum Leakage Acceptance Limits
Normal Condition
Single Fault Condition*
ECG
10
50
Universal
100
100
Earth
500
1000
*Single fault considered AC mains on applied part.