M Series Service Manual

43

20.33

On the Pressure Transducer
Simulator, set the pressure polarity
to positive, and select a pressure
of 500 mmHg.

On the unit, verify that the value displayed for the ART P2
pressure channel is in the range of 291 - 300 mmHg.
(Note: value will be displayed within 15 seconds.)

o

o

o

20.34

Connect the Pressure Transducer
Simulator to the first IBP channel
connector (P1). On the Pressure
Transducer Simulator, set the
pressure polarity to negative, and
select a pressure of -200 mmHg.

20.35

Zero the pressure channel.

The unit calibrates the pressure channel against a
negative pressure of -200 mmHg and displays the P1
ZEROED message.

o

o

o

20.36

On the Pressure Transducer
Simulator, select a negative
pressure of -250 mmHg.

On the unit, verify that the value displayed for the P1
pressure channel is in the range of -47 to -50 mmHg.
(Note: value will be displayed within 15 seconds.)

o

o

o

20.37

Connect the Pressure Transducer
Simulator to the second IBP
channel connector (P2). On the
Pressure Transducer Simulator, set
the pressure polarity to negative,
and select a pressure of -200
mmHg.

20.38

Zero the pressure channel.

The M Series CCT unit calibrates the pressure channel
against a negative pressure of -200 mmHg and displays
the P2 ZEROED message.

o

o

o

20.39

On the Pressure Transducer
Simulator, select a negative
pressure of -250 mmHg.

Verify that the value displayed for the P2 pressure
channel is in the range of -47 to -50 mmHg. Note: Value
will be displayed within 15 seconds.

o

o

o

Do this...

Verify that...

Pass/Fail/
NA