21.0 NIBP Transducer Calibration Test

The NIBP module’s pressure transducers are factory-calibrated prior to shipment. However, you can perform a two-
point calibration procedure periodically to ensure accurate pressure measurements.

This procedure is optional at 6 months, but should be performed annually or every 10,000 readings, whichever comes
first.

Tools Needed:

Fluke NIBP simulator (the values and procedure that we provide are specific to the BP Pump 2).

Test Setup:

Connect the Analyzer’s simulator hose to the NIBP connector on the M Series unit.

Configure the NIBP Analyzer to simulate cuff pressure. For example, on the Fluke Biomedical BP Pump:

Press the MODE button three (3) times to go into Tests mode.

Press the SELECT button once to access the Pressure Simulator screen.

These instructions apply to the Fluke Biomedical BP Pump; for equivalent devices, follow the manufacturer’s
instructions.

Make sure the ECG cable is not connected to the M Series unit.

If the SpO2 option is installed, make sure that the Masimo cable is NOT connected to the M Series unit.

Do this...

Verify that...

Pass/
Fail

21.1

Turn the Selector Switch to OFF.

After 10 seconds, press and hold
the fourth softkey from the left and
turn the Selector Switch to
MONITOR (for AED units, turn the
Selector switch to ON).

.

The M Series unit powers on in the NIBP Service Mode.

o

o

21.2

Press the NIBP Calib softkey.

The M Series unit displays the NIBP Transducer
Calibration Screen.

o

o

21.3

On the NIBP simulator set the
pressure parameter to 0 mmHg.

The NIBP Simulator displays a pressure reading of 0
mmHg.