23.0 NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer

The volume leak test verifies the integrity of the pneumatic system on the NIBP module. This test is optional at 6
months, but should be performed annually or every 10,000 readings, whichever comes first.

Tools Needed:

Fluke NIBP simulator (the values and procedure that we provide are specific to the BP Pump 2).

Test Setup:

Connect the Analyzer’s simulator hose to the NIBP connector on the M Series unit.

Configure the NIBP Analyzer for the volume leak test. For example, on the Fluke Biomedical BP Pump:

Press the MODE button three times to go into Tests mode.

Press the SELECT button twice to access the volume leak test.

Make sure the ECG cable is not connected to the M Series unit.

If the SpO2 option is installed, make sure that the Masimo cable is NOT connected to the M Series unit.

Do this...

Verify that...

Pass/
Fail

23.1

Turn the Selector Switch to OFF.

After 10 seconds, press and hold
the fourth softkey from the left and
turn the Selector Switch to
MONITOR (For AED units, turn
the selector switch to ON).

The M Series unit powers on in the NIBP Service Mode.

o

o

23.2

Press the Leak Test softkey.

The M Series unit displays the NIBP Leak Test Screen.

o

o

23.3

On the NIBP simulator, set the
pressure parameter to 200
mmHg.

The NIBP simulator displays a pressure reading of 200
mmHg.

o

o

23.4

On the M Series unit, press the
Close Valves
softkey.

The Valves status changes from OPEN to CLOSED.