M Series Service Manual

92

Step 1: Remove the High Voltage/Charger
Assembly from the main housing by lifting the
High Voltage/Charger Assembly upwards and
rotating it towards the back of the unit.

Step 2: Set the Assembly on the table. Pull apart
the High Voltage Module from the Charger
Assembly. These three components are referred to
as the “Module Cluster”.

Step 3: Depress the lever on the High Voltage
Cable to remove it from the mounting panel. Cut
the tie wrap attached to the mounting bracket that
secures the cables.

Step 4: (SpO

2

units only) To remove the SpO

2

module from the Isolated Power Supply, remove
the Mylar

®

tape. EtCO

2

module and power

supply are one assembly at the same location.

Charger assembly

High voltage
module

High voltage
capacitor

high voltage cable