M Series Service Manual
92
Step 1: Remove the High Voltage/Charger
Assembly from the main housing by lifting the
High Voltage/Charger Assembly upwards and
rotating it towards the back of the unit.
Step 2: Set the Assembly on the table. Pull apart
the High Voltage Module from the Charger
Assembly. These three components are referred to
as the “Module Cluster”.
Step 3: Depress the lever on the High Voltage
Cable to remove it from the mounting panel. Cut
the tie wrap attached to the mounting bracket that
secures the cables.
Step 4: (SpO
2
units only) To remove the SpO
2
module from the Isolated Power Supply, remove
the Mylar
®
tape. EtCO
2
module and power
supply are one assembly at the same location.
Charger assembly
High voltage
module
High voltage
capacitor
high voltage cable