0 shock test, M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 36

Advertising
background image

M Series Service Manual

26

13.0 Shock Test

Tools Needed:

QED 6 Defibrillator Analyzer, Stop Watch.

Test Setup:

Connect the universal cable via the adapter (D.N.I #3010-0378) to the defibrillator analyzer.

Ensure that a fully charged battery is installed in the unit.

NOTE If your M Series AED does not have manual override capability, do not perform this test.

Do this...

Verify that...

Pass/Fail/
NA

13.1

Turn the selector switch to
DEFIB mode.

(For AED units, turn the selector
switch to ON and select Manual
mode.)

13.2

Press the ENERGY SELECT
down arrow until 5J displays.

The unit displays DEFIB 5J SEL.

o

o

o

13.3

Press the CHARGE button. Wait
for the SHOCK button to
illuminate.

The unit displays DEFIB 5J RDY.

o

o

o

13.4

Press the SHOCK button.

The unit discharges 3J-5J into the simulator.

Note:

The displayed rhythm may change shape for 30
seconds before it returns to an original rhythm.
This is caused by the operation of the adaptive
bandwidth defibrillator recovery circuit.)

o

o

o

13.5

Press the ENERGY SELECT up
arrow until the unit displays 50J.

The unit displays the message, DEFIB 50J SEL.

o

o

o

13.6

Press the CHARGE button. Wait
for the SHOCK button to
illuminate.

The unit displays the message, DEFIB 50J RDY.

o

o

o

13.7

Press the SHOCK button.

The unit discharges 41J-55J (monophasic 43J-57J)
into the simulator.

o

o

o