Removing the printer/ recorder motor, Removing the printer/recorder motor – ZOLL M Series Defibrillator Rev R User Manual
Page 110
Advertising
M Series Service Manual
100
10. Removing
the
Printer/
Recorder
Motor
Tools Required
• #2 Phillips screwdriver
• Exacto-Knife
• Kapton tape
Setup
1. Remove the paper tray, pull out and press
up on the locking tab at the rear of the tray.
2. Remove the Front Panel Assembly.
3. Remove the Upper Housing Assembly.
4. Remove the System Board.
5. Remove the Battery Interconnect Board.
6. Remove the High Voltage Module
Assembly.
7. Remove the sensor, printhead and motor
cables.
8. Remove the System Interconnect Board.
To reinstall the Printer/Recorder Motor, reverse the steps.