M Series Service Manual

100

10. Removing

the
Printer/
Recorder
Motor

Tools Required

• #2 Phillips screwdriver

• Exacto-Knife

• Kapton tape

Setup

1. Remove the paper tray, pull out and press

up on the locking tab at the rear of the tray.

2. Remove the Front Panel Assembly.

3. Remove the Upper Housing Assembly.

4. Remove the System Board.

5. Remove the Battery Interconnect Board.

6. Remove the High Voltage Module

Assembly.

7. Remove the sensor, printhead and motor

cables.

8. Remove the System Interconnect Board.

To reinstall the Printer/Recorder Motor, reverse the steps.