M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 65

Advertising
background image

M Series Service Manual

55

V LEADS OFF message

displays.

If the user is not using V leads, attach V lead
connector terminator plug to the cable’s V lead
connector.

If a V1 lead wire metal snap comes in contact with
the patient’s skin, then the system will show all V
leads as OFF.

Remove V1 leads and others away from the
patient. Turn off the unit and wait ten seconds
before turning it back on.

CHECK PADS/POOR PAD

CONTACT message

displays.

Remove and reinsert PADS connector into the
universal cable.

Check for damaged defibrillator pads, wires and or
connector.

Check for dried out or expired defibrillator pads.

Clip (not shave) the patient’s hair and wipe pad
contact area dry.

Connect the cable to the test plug. The DEFIB PAD
SHORT message displays to indicate that the cable
is functioning properly.

If the DEFIB PAD SHORT message displays, then
check the connections of the pads to the patient and
to the defibrillator cable.

If the DEFIB PAD SHORT message does not
display, remove the defibrillator from service.

Connect universal cable to the
shorting plug. The DEFIB PAD
SHORT message should display,
when you SELECT PADS. If the
message does not display, then:

Try another universal cable.

Check the cable from the universal
cable connector to the High
Voltage Module.

Check the cable from the High
Voltage Module to the system
board.

Remove and replace the High
Voltage Module.

Remove and replace the system
board.

Call ZOLL Technical Support for
assistance.

Reported Problem

Recommended User Action

Recommended Technical Action