M Series Service Manual

69

PRESS CHARGE

Advisory message in conjunction with shock

advised.

PRESS SHOCK

Prompt issued in AED auto defib mode when defib is

charged (ready).

RECORDER FAULT 142

Strip chart system error.

Check paper tray and

paper path. Replace the

print head, system

interconnect board and

or the system board.

RECORDER FAULT 143

Strip chart failed power-up echo test.

Communications error.

Check paper tray and

paper path. Replace the

system interconnect

board and/or the system

board. Turn unit off and

back on again.

RECORDER FAULT 147

Strip chart printhead over temperature.

Check paper tray and

paper path. Replace the

print head, system

interconnect board and/

or the system board.

RELEASE BUTTONS

Simultaneous external paddle button presses

detected before unit reached full defib charge (ready

state).

Release buttons.

RELEASE SHOCK

Discharge switch(es) closed when pressing charge

button. Discharge button pressed before defib

reached ready state.

Release shock button.

Check paddles.

Replace controls board.

REPLACE BATTERY

Battery voltage is less than absolute minimum.

Shutdown imminent.

Replace with charged

battery.

Error Message

Explanation

User
Advisory

Technical Action