M Series Service Manual

39

20.0 IBP Monitoring Test

Tools Needed:

DNI Nevada 214B Patient Simulator with M Series interface cables, FOGG Transducer Simulator

Test Setup:

Set the BP output sensitivity scale to 5 uV/V/mmHg on the Patient Simulator

Connect the Patient Simulator to the IBP connector on the M Series CCT unit

Do this...

Verify that...

Pass/Fail/
NA

20.1

Set the selector switch on the M
Series CCT unit to MONITOR and
press the PARAM softkey on the
front panel to display the
Parameters menu.

Verify the parameter menu displayed.

o

o

o

20.2

Select the IBP option and press the
Enter softkey.

Verify the unit displays the IBP menu.

o

o

o

20.3

Press the Label softkey to display
the IBP Label menu.

Verify the unit displays the IBP label menu.

o

o

o

20.4

Select the Label P1 option and
press the Enter softkey.

Verify the M Series CCT unit displays the Label Options
menu.

o

o

o

20.5

Select the P1 option and press the
Enter softkey.

Verify the unit assigns the P1 label to the first pressure
channel (P1) and returns to the IBP menu.

o

o

o

20.6

Press the Label softkey to display
the IBP Label menu.

Verify the unit displays the IBP Label menu.

o

o

o

20.7

Select the Label P2 option and
press the Enter softkey.

Verify the M Series CCT unit display the Label Options
menu.

o

o

o

20.8

Select the ART option and press
the Enter sofkey.

Verify the unit assigns the ART label to the second
pressure channel (P2) and return to the IBP menu.

Verify that the M Series CCT unit displays advisory
messages to zero the pressure channels.

o

o

o

o

o

o