19.0 Temperature Test

Tools Needed:

IBP Temp Simulator 9100-0402-TF (or equivalent

1

)

Test Setup:

None

1. If you are using a Temperature Simulator that does have the same setting above, use the closest value available. Verify the

display temperature value is within 0.2 F or 0.1 C of the simulator setting

Do this...

Verify that...

Pass/Fail/
NA

19.1

Connect the 2-channel Y adapter
to the temperature connector
(T1/T2) on the M Series CCT
Unit.

19.2

Connectt the Temperature Probe
Simulator cables to the 2-channel
Y adapter.

19.3

On the Temperature Simulator,
set the T1 temperature channel
to 59.2 degrees F(15 degrees C).

Verify that the value displayed for the T1 temperature
channel betwwen 59.0 degrees to 59.4 degrees F (14.9
to 15.1 degrees C).

o

o

o

19.4

On the Temperature Simulator,
set the T1 temperature channel
to 112.8 degrees F (45 degrees
C).

Verify that the value displayed for the T1 temperature
channel betwwen 112.6 degrees to 112.8 degrees F
(44.9 to 45.1 degrees C).

o

o

o

19.5

On the Temperature Simulator,
set the T2 temperature channel
to 59.2 degrees F(15 degrees C).

Verify that the value displayed for the T2 temperature
channel betwwen 59.0 degrees to 59.4 degrees F(14.9
to 15.1 degrees C).

o

o

o

19.6

On the Temperature Simulator,
set the T2 temperature channel
to 112.8 degrees F (45 degrees
C).

Verify that the value displayed for the T2 temperature
channel betwwen 112.6 degrees to 112.8 degrees F
(44.9 to 45.1 degrees C).

o

o

o

Record your results on the Maintenance Test Checklist.