M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 38

M Series Service Manual
28
13.20
(Biphasic Units) Press the
CHARGE button and start timing
with a stopwatch. Stop timing
when the SHOCK button lights.
The charge time is between 3-6 seconds.
o
o
o
13.21
Press the SHOCK button.
On the strip chart, the Patient Current is between 20-24
Amps, and the Defib Impedance is between 46-54
Ohms.
Note: This will occur if configured to Auto-Generate
Strips on event.
o
o
o
13.22
(Monophasic Units) Press the
CHARGE button and start timing
with a stopwatch. Stop timing
when the SHOCK button
illuminates.
The charge time is between 4-8 seconds.
o
o
o
13.23
Press the SHOCK button.
On the strip chart, the Patient Current is between 20-24
Amps, and the Defib Impedance is between 46-54
Ohms.
Note: This will occur if configured to Auto-Generate
Strips on event.
o
o
o
13.24
(AED units only.) Disconnect the
universal cable from the
analyzer.
The unit will display CHECK PADS, and give anaudio
prompt.if configured for audio.
o
o
o
Record your results on the Maintenance Tests Checklist.
Do this...
Verify that...
Pass/Fail/
NA