M Series Service Manual

28

13.20

(Biphasic Units) Press the
CHARGE button and start timing
with a stopwatch. Stop timing
when the SHOCK button lights.

The charge time is between 3-6 seconds.

o

o

o

13.21

Press the SHOCK button.

On the strip chart, the Patient Current is between 20-24
Amps, and the Defib Impedance is between 46-54
Ohms.

Note: This will occur if configured to Auto-Generate
Strips on event.

o

o

o

13.22

(Monophasic Units) Press the
CHARGE button and start timing
with a stopwatch. Stop timing
when the SHOCK button
illuminates.

The charge time is between 4-8 seconds.

o

o

o

13.23

Press the SHOCK button.

On the strip chart, the Patient Current is between 20-24
Amps, and the Defib Impedance is between 46-54
Ohms.

Note: This will occur if configured to Auto-Generate
Strips on event.

o

o

o

13.24

(AED units only.) Disconnect the
universal cable from the
analyzer.

The unit will display CHECK PADS, and give anaudio
prompt.if configured for audio.

o

o

o

Record your results on the Maintenance Tests Checklist.

Do this...

Verify that...

Pass/Fail/
NA