M Series Service Manual

57

Displayed HR not accurate;

artifact present.

Reduce or eliminate ECG artifact due to
electrode or patient cable movement. Route
cables so that they don’t pull on electrodes or
swing excessively.

Ensure patient is motionless.

Check for possible excessive radio frequency
interference.

Verify a good connection of electrodes to the
patient.

Prepare the patient’s skin prior to the electrode
attachment.

Move patient cables away from other electrical
equipment, especially any RFI source.

Ensure ECG cable fits snugly in unit.

Change ECG cable.

Replace/reposition ECG electrodes.

Check for contamination on snaps.
Ensure springs are intact.

Check for intermittent ECG patient
cable or connector wiring.

Replace ECG input connector.

Replace ECG connector to the
system board cable.

Replace system board.

Wandering baseline.

See “Displayed HR not accurate.” above. Note that

in 90% of electrode issues, size and lead changes

don’t help.

Same as above example.

Electronic interference.

Check for possible excessive radio frequency

interference.

Move patient cables away from other electrical

equipment.

Turn off sources of excessive RFI.

Move M Series unit away from
RFI source.

Reported Problem

Recommended User Action

Recommended Technical Action