M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 42

M Series Service Manual
32
16.6
Increase the output to 140mA.
Output on the QED 6 analyzer is 140mA +/- 7 mA.
o
o
16.7
Check the pulse width.
Pulse width is 40mS +/-2mS.
o
o
16.8
Decrease the output to 60mA.
Decrease the rate to 30ppm.
Pacer rate on QED 6 analyzer is 29-31 ppm.
o
o
16.9
Increase the rate to 180ppm.
Pacer rate on QED 6 analyzer is 177-183 ppm.
o
o
16.10
Connect the ECG cable to the
M Series and QED 6 analyzer.
Select the ECG at 60 BPM on the
QED 6 analyzer.
The unit displays ECG at 60 BPM with no stimulus
markers.
o
o
16.11
Press the Async Pace softkey.
The unit displays ECG at 60 BPM with stimulus markers,
and displays the Async pace message.
o
o
16.12
Turn off the QED 6 analyzer. Set
Pacer Rate to 100ppm. Press the
RECORDER ON button.
The unit displays pace stimulus markers every 15mm +/-
1mm.
o
o
16.13
Press and hold 4:1 button.
The unit displays pace stimulus markers every 60 mm+/-
1.5 mm.
o
o
Record your results on the Maintenance Tests Checklist.
Do this...
Verify that...
Pass/
Fail