M Series Service Manual

32

16.6

Increase the output to 140mA.

Output on the QED 6 analyzer is 140mA +/- 7 mA.

o

o

16.7

Check the pulse width.

Pulse width is 40mS +/-2mS.

o

o

16.8

Decrease the output to 60mA.

Decrease the rate to 30ppm.

Pacer rate on QED 6 analyzer is 29-31 ppm.

o

o

16.9

Increase the rate to 180ppm.

Pacer rate on QED 6 analyzer is 177-183 ppm.

o

o

16.10

Connect the ECG cable to the
M Series and QED 6 analyzer.
Select the ECG at 60 BPM on the
QED 6 analyzer.

The unit displays ECG at 60 BPM with no stimulus
markers.

o

o

16.11

Press the Async Pace softkey.

The unit displays ECG at 60 BPM with stimulus markers,
and displays the Async pace message.

o

o

16.12

Turn off the QED 6 analyzer. Set
Pacer Rate to 100ppm. Press the
RECORDER ON button.

The unit displays pace stimulus markers every 15mm +/-
1mm.

o

o

16.13

Press and hold 4:1 button.

The unit displays pace stimulus markers every 60 mm+/-
1.5 mm.

o

o

Record your results on the Maintenance Tests Checklist.