M Series Service Manual

70

REPLACE EtCO

2

SENSOR

EtC0

2

SENSOR WARM UP message displays for

more than five minutes. Sensor defective.

Replace sensor cable.

REPORT FULL

Summary report memory full.

Erase summary report.

REPORT HALTED

Summary report stops printing unexpectedly.

Turn unit off and then

back on again. Print

Summary again. If fault

persists, replace system

board.

RESERVED 1

The watchdog timer is not functional in the unit.

Turn off unit and then

turn on to reset. If fault

persists, replace system

board.

RETRY ANALYSIS

Advisory message in conjunction with noisy ECG.

Analysis halted.

SELECT 30J FOR TEST

Attempt to run a self test at an energy other than 30J.

SELECT DEFIB MODE

Analyze button pressed in pace or monitor mode.

SELECT LIMB LEADS

Paddles or augmented ECG leads selected when

continuous analysis active or started.

Select limb leads I, II,

III or MFE

SELECT PADS

Lead I, II, or III selected when analyze pressed.

SET CLOCK

Real time clock failure: invalid date or time.

Set date and time and/

or replace system

board.

SET PACE mA

Multiple copy errors are the product of intended

software or memory errors. If error reoccurs other

than on entering pace the first time or after more than

10 minutes in other mode, the unit could be broken.

Set pace current. If

broken, replace system

board.

Error Message

Explanation

User
Advisory

Technical Action