M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 71

Advertising
background image

M Series Service Manual

61

CARD FULL

Memory Card Full.

CHECK CO

2

SENSOR

EtCO

2

Sensor is unplugged or defective.

Check that sensor cable

is plugged in and seated

properly. Check that

sensor is not exposed to

excessive heat. If

problem persists,

replace the sensor.

CHECK CO

2

ADAPTER

Airway adapter is removed, occluded or adapter

zeroing needs to be performed or was performed

incorrectly.

Replace/Clean airway

adapter. Zeroing

performed

automatically.

CHECK MEMORY CARD

No card detected during manual or semi-automated

modes.

CHECK PADS

Message displayed in conjunction with either POOR

PAD CONTACT or DEFIB PAD SHORT.

Ensure pads are

coupled to patient.

Check /replace pads

and universal cable.

Replace system board.

CHECK PATIENT

Background ECG analysis detects shockable rhythm.

CHECK PULSE

Alternate message for NO SHOCK ADVISED

message.Message also shown after delivering third

shock when auto analyze 3 times option is enabled.

CHECK RECORDER

Produced when paper tray is empty, paper jams or

recorder door is opened.

Replace paper sensor

board, system

interconnect board,

and/or system board.

Error Message

Explanation

User
Advisory

Technical Action