0 heart rate display test, M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 28

Advertising
background image

M Series Service Manual

18

7.0 Heart Rate Display Test

Tools Needed:

• ECG Cable (3, 5 or 12 leads).

• QED 6 Defibrillator Analyzer.

Test Setup:

Turn the selector switch to MONITOR (For AED units, turn the selector switch to ON and select Manual override.)
Press LEAD button until “I” displays.

Connect the ECG leads to the QED 6 Defibrillator Analyzer.

Connect the ECG cable to the unit.

Do this...

Verify that...

Pass/Fail/
NA

7.1

Set the QED 6 Defibrillator
Analyzer

t

o 120BPM.

The Heart Rate displays as 120 +/- 2 bpm.

o

o

o

Record your results on the Maintenance Tests Checklist.