M Series Service Manual

72

SYSTEM FAULT 5

ECU RAM test failure or ROM checksum test

failure.

Turn off unit and then

turn on to reset. If fault

persists, replace system

board.

SYSTEM FAULT 6

No communications received from ECU for 4

seconds.

Turn off unit and then

turn on to reset. If fault

persists, replace system

board.

SYSTEM FAULT 7

The A/D converter is not performing conversions in

a timely manner.

Replace system board.

SYSTEM FAULT 36

P1MON is less than 412 counts or greater than 612

A/D counts. Pace/defib is disabled as long as

condition exits.

Replace system board.

SYSTEM FAULT 37

Disable pace/defib and MFE monitoring.

Replace system board.

SYSTEM FAULT 38

Failure to shutdown after “shutdown order” is

written to the RTC.

Replace system board.

TEST FAILED

MCU performed ipeak test (defib peak current) and

unit failed during 30J self test.

Replace universal

cable, paddles or high

voltage module,

capacitor, or system

board.

TEST OK

MCU performed ipeak test (defib peak current) and

unit passed 30J self test.

USE PADDLE DISCHG

Front Panel discharge button is pressed when either

external paddles or internal spoons with discharge

buttons are connected.

Error Message

Explanation

User
Advisory

Technical Action