If you are using the Fluke® BiomedicalCuftllnk Simulator,the volume leak reading should be <10.

6

If you are using the Fluke® BiomedicalCuftllnk Simulator,the volume leak reading for a failure should be >10.

23.5

Press the START TEST softkey
within 30 seconds after closing
the valve.

Note:

This test takes
approximately 3 minutes.

After approximately 1 minutes, a number appears in the
upper middle area of the NIBP simulator display.

If the simulator:

Displays a Volume Leak reading <5,

5

then the M

Series unit has passed the test.

Displays a Volume Leak reading >5,

6

then the M

Series unit has failed the test.

Displays no Volume Leak reading, but maintains a
stable pressure reading at or above 200 mmHg, then
the M Series unit has passed the test; there is no
volume leak.

In addition, the M Series unit displays the simulator’s
pressure reading in the “Cuff Pressure” field.

After approximately 3 minutes, the valves open on the M

Series unit.

o

o

23.6

On the NIBP Analyzer, press the
STOP TEST softkey.

The NIBP simulator terminates the Volume Leak Test.

o

o

23.7

On the M Series unit, press the
EXIT softkey twice.

The M Series unit returns to the main NIBP Service
Mode screen, then to normal Monitor mode operation.

o

o

Record your results on the Maintenance Tests Checklist.

Do this...

Verify that...

Pass/
Fail