0 defibrillator self test, Warning! shock hazard – ZOLL M Series Defibrillator Rev R User Manual

Page 33

Advertising
background image

M Series Service Manual

23

11.0 Defibrillator Self Test

WARNING!
SHOCK HAZARD!

TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY
BEFORE YOU START CONDUCTING THE DEFIBRILLATOR TESTS.

Keep hands and all other objects clear of the multi-function cable connections and defibrillator
analyzer when discharging the defibrillator.

Before you discharge the defibrillator, warn everyone near the equipment to STAND CLEAR.

CAUTION

Do NOT internally discharge the unit more than 3 times in 1 minute. Note that the rapid repetition of internal discharges at

more than 30 Joules may damage the unit.

Tools Needed:

• MFC Test Port Connector 1004-0053-99 with universal cable.

• MFC Test Adaptor Connector (Fluke Biomedical Part Number 3010-0378 or equivalent).

• QED 6 Defibrillator Analyzer.

• ECG Cable.

• Stop watch.

Test Setup:

Ensure the unit is turned off

Ensure the ECG cable is connected to the unit and analyzer.