0 notch filter test, M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 30
Advertising
M Series Service Manual
20
9.0 Notch Filter Test
Tools Needed:
QED 6 Defibrillator Analyzer.
Test Setup:
Connect the ECG cable to the QED 6 Defibrillator Analyzer.
Connect the ECG cable to the unit
.
Do this...
Verify that...
Pass/Fail/
NA
9.1
Turn the selector switch to
MONITOR mode.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
9.2
Select lead I, size 3x.
Select 60Hz (or 50 Hz for a 50Hz
unit) on
the QED 6 Defibrillator Analyzer.
9.3
Press RECORDER button.
Verify that the waveform amplitude on the strip chart is
less than 1.5 mm.
o
o
o
9.4
Press RECORDER button to
stop printing.
Record your results on the Maintenance Tests Checklist.