Removing the high voltage capacitor assembly – ZOLL M Series Defibrillator Rev R User Manual

Page 106

Advertising
background image

M Series Service Manual

96

8. Removing

the High
Voltage
Capacitor
Assembly

Tools Required

• #2 Phillips screwdriver

• Exacto-Knife

• Mylar

tape

Setup

1. Remove the Front Panel Assembly.

2. Remove the Upper Housing Assembly.

3. Remove the System Board.

4. Remove the Battery Interconnect Board.

5. Remove the High Voltage Charger

Assembly.

6. Lift the Capacitor Assembly upwards from

the chassis. (The Capacitor Assembly is still
connected to the High Voltage Module
Assembly.)

To reinstall the High Voltage Capacitor Assembly, reverse the steps.

WARNING! This unit may contain lethal voltages. You MUST completely discharge the high voltage

capacitor before removing from unit. DO NOT SHORT THE TERMINAL ENDS OF THE
CAPACITOR.

Step 1: Open the High Voltage Capacitor plastic
isolator by lifting the

Mylar

tape

.

Step 2: Bleed the excess voltage using a resistor
with values of approximately 5 kohms, 25 watts
for 10-20 seconds. Measure the voltage on the
capacitor terminals.

Step 3: When the capacitor voltage is at 0 VDC,
disconnect the inductor wires and capacitor
terminal wires.