93

Step 5: Lift the SpO

2

module straight up from the

foam and disconnect the ribbon cable from the
Isolated Power Supply.

Step 6: (Biphasic units only) Disconnect the
biphasic defibrillator capacitor from the rear
panel connector.

Step 7: (Biphasic units only) Disconnect the
Molex connector from the biphasic module.

Step 8: (EtCO

2

units only) Remove the EtCO

2

connector by gently lifting the connector away
from the back panel. To remove the EtCO

2

and

isolated power supply, refer to Step 5.