M Series Service Manual

37

18.4

On the Novametrix simulator,
set% CO

2

to 5%.

The EtCO

2

reading of 34-42mmHg displays on the M

Series monitor. The Respiration Rate (RR) of 22-24
displays on the M Series monitor.

Note that you may need to wait up to 10 seconds for the
unit to stabilize.

o

o

18.5

Press WAVE 2 softkey

For CCT units: Press the Trace
softkey, select SET TRACE 2,
and press Enter. Select EtCO2,
press Enter.

Verify the EtCO

2

waveform displays.

o

o

18.6

Press the RECORDER button.

Verify the EtCO

2

waveform prints

Note that the CO

2

waveform is displayed and printed at

12.5 millimeters per second scale.

o

o

Record your results on the Maintenance Tests Checklist.

Do this...

Verify that...

Pass/
Fail