Zoll m series error messages, M series service manual – ZOLL M Series Defibrillator Rev R User Manual

Page 68

Advertising
background image

M Series Service Manual

58

Zoll M Series Error Messages

The following is a list of Zoll M Series error messages that may appear on your display. The “User Advisory” column
informs you about an action in progress or provides feedback on a user correctable situation that typically does not
require further technical support. The “Technical Action” column describes what you as a technician can do to correct
the situation. Note that these messages will sometimes overlap part of the waveform display.

First, attempt to clear the message by turning the Selector Switch to OFF for ten seconds, then back to the desired
operating mode. If the fault persists, call ZOLL Technical Service.

Error Message

Explanation

User
Advisory

Technical Action

200J MAX BIPHASIC

User attempted to set defibrillation energy >200J on

Biphasic Unit. No higher energy is available.

50J MAX

Energy < 50J for internal paddles. No higher energy

is available.

ADJUST ECG

Unit is in sync mode and heart rate is < 20 BPM.

Or, QRS size set too small for proper

synchronization.

ANALYSIS HALTED

ECG analysis halted due to user interaction such
as:

Lead/size change

Analyze button was pressed again

Impedance fault

Charging error detected in auto defib mode

ANALYSIS RESTARTED

This is a user prompt issued simultaneously with

ECG TOO LARGE or ECG TOO SMALL. Device

detected ECG signals out of range, automatically

adjusted ECG size and is now restarting its

shockable rhythm analysis sequence.