M Series Service Manual

12

Do this...

Verify that...

Pass/
Fall

4.1

Turn the selector switch to
MONITOR.

(For AED units, turn the selector
switch to ON and select Manual
mode.)

The unit does not turn on.

o

o

4.2

Turn the selector switch to the
OFF position.

4.3

Adjust the power supply voltage to
10.3V and turn the selector switch
to MONITOR (for AED units, turn
the selector switch to the ON
position).

The unit turns on.

o

o

4.4

Low Battery Test

Set voltage to 9.8V.

The unit does not display the LOW BATTERY message.

o

o

4.5

Set voltage to 9.4V.

The unit displays the LOW BATTERY message within 30
seconds.

o

o

4.6

Shut Down Voltage Test

Set voltage to 8.5V.

The unit shuts off within 30 seconds.

o

o

Record your results on the Maintenance Tests Checklist.

?