medSim 300B

Operators Manual

4-4

ART = 120/80

ART=Arterial, LV=Left Ventricular

LV = 120/0

eic153.eps

Figure 4-1. medSim 300B Simulated Waveforms for Normal Blood Pressure Relationships across the

Aortic Valve

LV = 180/0 (+50%)

LV = 144/0 (+20%)

LV = 138/0 (+15%)

LV = 132/0 (+10%)

LV = 126/0 (+5%)

Art = 120/80

eic154.eps

Figure 4-2. medSim 300B Simulated Systolic Pressure Gradients Indicating a Defective Aortic Valve

Note

Increases in left ventricular (LV) pressure over arterial (ART) pressure

simulate a defective aortic valve.

RV = 25/0

PA = 25/10

PA=Pulmonary arterial,RV=Right Ventricular

eic155.eps

Figure 4-3. medSim 300B Simulated Waveforms for Normal Blood Pressure Relationships across the

Pulmonary Valve