Fluke Biomedical medSim 300B User Manual
Page 102
Advertising
medSim 300B
Operators Manual
4-4
ART = 120/80
ART=Arterial, LV=Left Ventricular
LV = 120/0
eic153.eps
Figure 4-1. medSim 300B Simulated Waveforms for Normal Blood Pressure Relationships across the
Aortic Valve
LV = 180/0 (+50%)
LV = 144/0 (+20%)
LV = 138/0 (+15%)
LV = 132/0 (+10%)
LV = 126/0 (+5%)
Art = 120/80
eic154.eps
Figure 4-2. medSim 300B Simulated Systolic Pressure Gradients Indicating a Defective Aortic Valve
Note
Increases in left ventricular (LV) pressure over arterial (ART) pressure
simulate a defective aortic valve.
RV = 25/0
PA = 25/10
PA=Pulmonary arterial,RV=Right Ventricular
eic155.eps
Figure 4-3. medSim 300B Simulated Waveforms for Normal Blood Pressure Relationships across the
Pulmonary Valve